Table
of Contents Back | Next | Home |
![]() Missoula Technology & Development Center |
Evaluation of Optical Instruments for Real-Time Continuous Monitoring of Smoke Particulates
Two basic types of instruments were used in the evaluation: optical real-time continuous particulate monitors and gravimetric instruments. Tables 1 and 2 summarize the specifications and accessories for the real-time instruments.
Price (dollars) |
Size (inches) |
Weight (pounds) |
Mass concentration range (µg/m³) |
Power requirements |
Internal calibration |
Internal data storage |
Internal battery |
Analytical filter | |
---|---|---|---|---|---|---|---|---|---|
Met One GT-640 | 5,382 | 15.7x11.7x8 | 23 | 0 to 10,000 | 12 V dc, 1 A | X | X | ||
MIE DataRam | 11,000 | 5.28x7.25x13.63 | 11.7 | 0 to 400,000 | 6 V dc, 3 A | X | X | X | X |
Radiance Research | 4,795 | 22x5x7 | 5.7 | 0 to 1,000 | 12 V dc, 1 A | X | |||
Optec NGN-3 | 12,900 | 10.7x8.2x16.5 | 27 | 0 to 16,666 | 13.8 V dc, 4.5 A | ¹X | |||
Andersen aethalometer | 16,645 | 19x12.5x10.5 | 20 | 0 to 1,000 | Lin power, 115 V ac | X | ²X | ||
¹ Requires external supply of span gas. | |||||||||
² Data storage with 3.5-in floppy disk installed. |
Inlet heater (standard) |
PM2.5 cutoff |
Environmental enclosure (standard*) |
Relative humidity (standard**) |
Ambient temperature |
Meteorological equipment | |
---|---|---|---|---|---|---|
Met One GT-640 | Standard | X | Standard* | Standard** | X | X |
MIE DataRam | X | X | ||||
Radiance Research | X | X | X | |||
Optec NGN-3 | Standard | X | Standard* | |||
Andersen aethalometer | X | X | ||||
* Self-contained fiberglass enclosure. | ||||||
** Internal relative humidity sensor standard, ambient relative humidity sensor optional. | ||||||
Wind sensor, ambient temperature, ambient relative humidity. | ||||||
Continuous heater or proportional, selectable heater. | ||||||
* Internal temperature standard, external temperature optional. |
![]() ![]() |
Back | Next Table of Contents |
![]() Missoula Technology & Development Center |