Evaluation of Optical Instruments for Real-Time Continuous Monitoring of Smoke Particulates
1998 Laboratory TestsThe Met One GT-640 was not available for these tests.
Field TestsOnly one GT-640 was available for the field tests. Results (figure 13) show the instrument overestimated the mass concentrations by 10 percent in the field environment (slope = 1.10) when compared to the FRM. The R² value was poor at 0.35 but only 12 data points were obtained.

Figure 13Comparison of the Met One GT-640 and gravimetric
results from the 1999 field tests.